specifications

The 4G GPTAS is presently a conventional neutron triple-axis spectrometer. The basic specifications are given in the table below.

MonochromatorRotating-shield type
ReflectionsPyrolytic graphite 002 reflections
Size120mm(H)X170mm(V)
FocusingVertical focusing
Scattering angle rnage18 < 2th_M < 47.4 degrees
k_i range5.99 < k_i < 2.33 Ang-1
Wavelength range1.1 < lambda_i < 2.7 Ang
FilterInpile PG filter (removable)
Sample tableAir floating
Goniometer1 ton Goniometer
Tilt and TranslationMotor controlled
Beam size40mm(H)X60mm(V) at maximum
Scattering angle range-2 < 2th_S < 119 degrees
AnalyzerAir floating
ReflectionsPyrolytic graphite 002 and 004 reflections
Size190mm(H)X100mm(V)
FocusingVertical focusing
FilterPG filter before analyzer (removable by hand)
Collimators#1 10′, 20′, 40′
#2 10′, 20′, 40′, 80′
#3 10′, 20′, 40′, 80′
#4 10′, 20′, 40′, 80′
Control Softwareunified filman
Temperature and magnetic field controllable
Sample environment
Low temperatureGM (2.4 K~300 K; default)
Orange cryo (1.5 K~300 K)
1 K cryo (0.7 K~100 K)
3He cryo (0.3 K~100 K)
Dilution fridge (0.05 K~4 K): liquid He required
High-temperatureHigh-temperature CTI
Magnet1 T electric field magnet (2.4 K~ 300 K)
5 T vertical superconducting field magnet (1.5 K~300 K)
5 T horizontal superconducting field magnet (1.5 K~300 K)

Please ask instrument scientists for availability if you would like to use the sample enviroment other than GM.